Overview
The Algon NI 330 is an infant incubator designed to provide a controlled thermal environment for newborns. It combines air and skin temperature control with servo-controlled humidity, a double-wall enclosure and an air curtain to help maintain stable conditions during care.
Optional oxygen control, SpO₂ monitoring, an integrated weighing scale and a height-adjustable stand allow the system to be configured for different neonatal care requirements.
Features
- PID temperature regulation: Continuously adjusts heating output based on the difference between the set and measured temperatures.
- Air and skin temperature control: Supports both control modes, with two temperature-probe channels for simultaneous monitoring.
- Uniform air circulation: Advanced air circuit design provides temperature uniformity of less than 0.5°C.
- Double-wall enclosure and air curtain: Help reduce heat loss when the access panel is opened.
- Servo-controlled humidity: Provides humidity adjustment and monitoring.
- Comprehensive alarm system: More than 30 alarm types.
- Dual heater protection: Incorporates a thermocouple and a mechanical temperature switch.
- Trend review: Supports 2-, 4-, 8-, 12- and 24-hour trends, with records extending up to seven days.
- Adjustable mattress: Mattress tilt of ±12°.
- Accessible storage: Rotatable storage drawers provide access from different directions.
- Optional functions: Touchscreen, lifting stand, weighing scale, oxygen monitoring or servo control, SpO₂ monitoring, X-ray cassette and oxygen cylinder holder.
- Additional catalogue-listed accessories: An adjustable observation lamp with three intensity levels and LED phototherapy equipment; inclusion should be confirmed for the selected configuration.
Technical Parameters
| Category | Parameter | Specification |
|---|---|---|
| Temperature control | Control modes | Air and skin temperature control |
| Air temperature setting range | 20–37°C | |
| Air temperature override range | 37–39°C | |
| Skin temperature setting range | 34–37°C | |
| Skin temperature override range | 37–38°C | |
| Thermal performance | Temperature uniformity | <0.5°C |
| Temperature variability | <0.3°C | |
| Warm-up time | <35 minutes, from 22°C at 50% RH to 33°C | |
| Air velocity across mattress | <10 cm/s | |
| Operating noise inside hood | ≤45 dBA | |
| Humidity control | Control range | 30–95% RH |
| Display and control accuracy | ±3% RH | |
| Water reservoir capacity | 1,500 mL | |
| Oxygen servo control — optional | Control range | 21–65% |
| Display accuracy | ±2% of full scale | |
| Display resolution | 1% | |
| Weighing scale — optional | Measurement range | 300 g–8 kg |
| Display resolution | 1 g | |
| Accuracy | ±5 g | |
| Mattress | Tilt angle | ±12° |
| Data review | Trend intervals | 2, 4, 8, 12 and 24 hours; up to seven days |
| Storage and shipping | Temperature | −25 to +60°C |
| Humidity | 0–95% RH |