Overview
Radiant Infant Warmer AG-RW-4000, with its comprehensive range of features, represents a significant advancement in newborn care technology. Its combination of innovative design, intuitive operation, and medical functionality creates a safer care platform that enhances both patient outcomes and user experience. Healthcare providers can trust the AG-RW-4000 to deliver the highest level of care for every infant, ensuring a warm, safe start to life.
Features
- 10.4″ LCD touch screen
500W Far Infrared Heater - Baby, Manual and Pre-heating mode
- Hand-Free Alarms Silence Sensor
- Environment light and noise level monitoring
- Transparent Guard Door with Damper
- Tilting Trenddelenburg ±13°
- Examination Lamp (0-1500lux)
- IV-Pole and Shelf
- Storage Drawer
- Auxiliary Power Output *2
Technical Parameters
| Powerrequirements | Customer specified |
| Heater capacity | 500w |
| Canopy mobility range | Horizontal rotation 320(1 60bothclockwise and counterclockwise) Radiation range ±30° |
| Temperature control | Servo (Baby) Mode/ Manual/Preheating mode |
| Skin temperature | Setting range: 34.0~38.0℃ |
| Accuracy:±0.3° | |
| Display range:20.0~42.0C | |
| Heater output | Setting range:0~100% (in 5% increments) |
| ndicatlonrange:0~100% (in 1% increments) | |
| Timer | APGAR mode/CPR mode switchable |
| APGAR Beep: 1,3 (option),5,10 min | |
| Examinationlighting | Power: 6W LED (2W *3) |
| LED lamp Adjustable illumination intensity:0~1500lux | |
| LCD display | 10.4″ LCD touch screen |
| Pixel: 800*600 | |
| Bedtray | Inclination angle (back and forth):±13° |
| Mattress surface: 500 * 700*25mm | |
| Mattress surface to ground height (standard): 90cm | |
| Guard door: damping system | |
| Cable access hole: 3 | |
| Dimensions | 84(W)*110(D)*186(H) cm (standard) |
| Net Weight | 90kg |
| Alarmssilence | Hand free |
| Height adjustment | Up to 30cm |
| Weight Scale | |
| Weightrange | 300 g to 8000 g |
| Weight display resolution | 1g |
| Weight accuracy | ±10g |
| X-ray Cassette | |
| Dimension | 33*35cm |
| Drawer | |
| Weight limit | 5kg |
| Pulse Oximeter Unit | |
| Display range | 1~100% SpO2 |
| Displayaccuracy | ±2%(70~99%) |
| Pulseratedisplay range | 30~240bpm |
| Pulse rate dis play acuracy | 1%+1bpm |
| Resuscitation | |
| Manometer | Pressure indication: -2~10KPa(-20~100cmH2O) |
| Resusci-flow unit | PMAX setting range:≤6KPa (60cmH2O) |
| ‘ Suction Unit | |
| Negativepressurerange | Suction pressure control range:-24~0 KPa(-180~0 mmHg) |
| Air pumpingefficiency | 0~15L/min |
| Storage/Shipping Information | |
| Temperature | -25°~60℃ |
| Humidity | 0~95% RH |